Publications

Gravimetric Wear Analysis and Particulate Characterization of Bilateral Facet-Augmentation System – PercuDynTM)

Publication Information:

Bhattacharya, S., A. Nayak, V.K. Goel, C. Warren, S. Schlaegle, L. Ferrara, “Gravimetric Wear Analysis and Particulate Characterization of Bilateral Facet-Augmentation System – PercuDyn™,” Bio-Medical Materials and Engineering, Vol. 20, Number 6 , pp. 329-338, January 2011.

Year: 2011

Topics: Particle Characterization

Dynamic stabilization systems are emerging as an alternative to fusion instrumentation. However, cyclic loading and micro-motion at various interfaces may produce wear debris leading to adverse tissue reactions such as osteolysis. Ten million cycles of wear test was performed for PercuDyn™ in axial rotation and the wear profile and the wear rate was mapped. A validation study was undertaken to assess the efficiency of wear debris collection which accounted for experimental errors. The mean wear debris measured at the end of 10 million cycles was 4.01 mg, based on the worst-case recovery rate of 68.2%. Approximately 40% of the particulates were ellipsoidal and the remaining particles were of irregular shapes. The PercuDyn™ exhibited an average polymeric wear rate of 0.4 mg/million cycles; substantially less than the literature derived studies for other motion preservation devices like the Bryan disc and Charité disc. Wear debris size and shape were also similar to these devices.